Item
|
Content
|
Technical Parameter
|
Power Supply
|
Voltage
|
220V
|
Frequency
|
50Hz
|
Capacity
|
≥40kVA
|
internal resistor
|
≤0.15Ω
|
Generator
|
|
Power Output
|
32KW
|
Inverter Frequency
|
30 KHz
|
photography
|
Tube Voltage
|
40kV—150kV
|
Tube Current
|
10mA—320mA
|
Exposure time
|
1.0ms—6300ms
|
Digital Controlled
X-ray Tube
|
Tube Focus:Large Focus/Small Focus
|
1.2mm /0.6mm
|
Input Power
|
Big focus 40kW / Small Focus 20kW
|
thermal capacity
|
110KJ
|
Rotary anode speed
|
2800rpm
|
Radiography Table
|
Table longitudinal movement
|
≥900mm
|
Table transverse movement
|
≥220mm
|
Pillar movement along table
|
≥1300mm
|
Detecting holder movement
|
≥500mm
|
Tube
assembly up-down movement
|
500mm~1280mm
|
Collimator
|
Manual Multi-leaf
|
Table use fixed grid
|
Grid density 103L/INCH, grid 10:1 SID: 120cm, fixed model: 15″×18″
|
Bucky Stand
|
Radio-graphic device movement along
pillar
|
≥1300mm
|
Cassette SID:
|
450 ~1780mm
|
Fixed grid:
|
Grid density 103L/INCH, grid 10:1 ;SID 180cm;Size: 15″×18″
|
Flat Panel Detector
|
Vision size: 358mm(H)×430 mm(V)
Pixel Matrix: 2560(H)×3072(V)
Pixel gap:139 μm
Imaging time:less than 10s
Limiting resolution:3.5 lp/mm typ
A / D transform:14 bit
Energy range:40 - 150 kVp
Biggest input dose(low transmission gain): 4 mR / frame
Date out:A 16-bit digital output Ethernet(1000BASE - T)
Command control:Ethernet(1000BASE - T)
Power input: DC 24V 4.8A
|
Work Station
|
CPU System
|
Brand: DELL OPTIPLEX 7010 commercial use
Processor:Core™ i5-3570 processor
RAM:4GB DDR3
Hardware:500GB SATA (7200 rpm)
CD-driver:suit MT’s 16X variable-speed DVD + / - RW With functions of double written
Network Card: Broadcom NetXtreme
10/100/1000 PCIe Gigabit LAN GNTB-A
, High display card: 1GB AMD RADEON HD
7570,FH,with DVI-VGA Adapter
Slop: PCIE seriel port and parallel port,
all height, MT
|
Color LCD Monitor
|
19 inches 1M LED backing light LCD
display, revolution1280x1024、5:4screen、dot pitch:0.294mm, contrast ratio: 1000:1, brightness: 250cd/㎡, Grey feedback time:5ms 、visual Angle: 160/170°
|
Monochrome Monitor(Optional)
|
1280*1024,19inches,brightness
1000cd/m2, Contrast Ratio 900﹕1
|
Work Station Software
|
Basic operation: change control console password, edit ID, acquisite images.
|
Additional operation:Add new check, edit present checking info, add
new position, change image aquisition order, multi-checking agreement chance,
manual adjustment of exposure parameters, automatic exposure control mode,
focus choice, patient body-type choice, tube capacity check, ESA curve
choice, image cutting, note added (sent to DICOM workstation) , mark on
images, rotate or overturn, full-size image observation, check patient info
and dose info, accept or refuse images.
|
Image management: change order, patient basic info editing, inquiry
history images, resend history images, re-print history images, check images
mark info, review history images, manage refused images, space reclaimed,
image protection, and manual image deletion etc.
|
System management:ID
edition, change ID password, ED refrigeration set, statistics info checking, detector calibration, equipment control, output order